Use the rolodex controls or search box below to find tests. You can enter a test name, CPT code, test code, or alias in the search box.

Test Name: Calcium, Ionized, Serum
Test Code: 004804
Other Names: N/A
Test Includes: Calcium, Ionized, Serum
Specimen Requirements: Serum, 1 mL
Container: Gel-barrier tube
Collection Instructions: Centrifuge and send unopened gel-barrier tube with vacuum intact. It is critical that the vacuum not be disrupted. Exposure of the serum to room air causes a loss of carbon dioxide to the atmosphere. This changes the sample pH and can prevent the lab from measuring the ionized calcium. To ensure that testing can be performed, send a separate tube for ionized calcium and adhere a piece of labeling tape over the top of the tube with the following written on it "IONIZED CALCIUM-DO NOT OPEN."
Stability: Do NOT freeze.
Rejection Criteria / Special Instructions: Centrifuge and send unopened gel-barrier tube with vacuum intact. It is critical that the vacuum not be disrupted. Exposure of the serum to room air causes a loss of carbon dioxide to the atmosphere. This changes the sample pH and can prevent the lab fro
Methodology: Ion-selective electrode (ISE)